Shenzhen HAD Trading Co., Ltd.
|
Place of Origin: | Zhejiang, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Adjustable Veterinary Infusion Pump With Button Operation History Records
Product advantages
Technical parameters
Bolus rate
|
100~1000ml/h |
Applicable infusion sets |
All brands of standard infusion sets |
Injection accuracy |
±3% |
Infusion rate range |
1~2000ml/h |
Infusion increment |
0.1ml/h |
Rate mode |
1~2000ml/h |
Selectable modes |
Rate mode, Drip mode,Time mode, Body weight mode |
KVO rate |
Adjustable, 1ml/h~5.0ml/h |
Time mode |
Volume limit: 0.1~9999.9ml |
Time: 1~6000 minutes |
|
Panel instructions |
1, Infusion rate 2, Infusion preset volume 3,Accumulated infusion volume 4, Battery capacity 5, Battery charging indication 6,AC power supply connection indication 7,Over 8, OCCL. Alarm 9, Air 10, Bed 11,Pressure Value 12, Flow rate , etc. |
Battery
|
Rechargeable lithium polymer battery 7.4V, 3000mA, working for 4 hours(25ml/h) after being fully charged |
Body weight mode |
Weight: 0.1~300kg Volume limit: 0.1~999.9ml Drug limit: 0.1~999.9mg Dose:0.1~9999.9mg Unit: mg/kg/h, ug/kg/min |
Accumulated volume |
0.1~9999.9ml |
Preset range |
0.1~9999.9ml |
Power supply |
AC100240V 50/60HZ |
Classification |
Class II, type BF |
Occlusion value |
39.2kPa~137.2kPa, adjustable, pressure show on the graphic |
Alarm function |
Air, over, OCCLU, out of battery, low battery, CPU communication error, drop error, specification error, no operation remind, door open, no AC power supply, installation error. |
Electrical safety |
Meet the requirements of EN60601-1:1990+A1:1993+A2:1995 |
Storage condition |
Environment temperature: -20+55, air pressure (50106)kPa,relative humidity95% |
Max power consumption |
25VA |
EMC |
Meet the requirements of IEC60601-1-2(2007 version) |
Operation condition |
Environment temperature: 5+40, air pressure (86106) kPa, relative humidity80% |
Overall maximum size |
125*140*205mm (Length*Width*Height) |
Gross weight |
2.8kg |
Vet infusion pump