Shenzhen HAD Trading Co., Ltd.
|
Place of Origin: | Zhejiang, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Colorful LCD Screen Veterinary Syringe Pump For Drug library and factory
Product advantages
Technical parameters
Applicable syringes
|
All brands of standard syringes: 5, 10, 20, 30 and 50 (60) ml |
Bolus rate |
50ml:400-1500ml/h 30ml: 400-600ml/h 20ml: 400ml/h 10ml: 200ml/h 5ml: 100ml/h
|
Injection accuracy |
±2% |
Injection rate range |
5ml: 0.1ml/h---100ml/h 10ml: 0.1ml/h---200ml/h 20ml: 0.1ml/h---400ml/h 30ml: 0.1ml/h---600ml/h 50(60)ml: 0.1ml/h---1500ml/h |
Injection increment |
0.1ml/h |
Drug names |
More than 20 drug names, such as Penicillin, Vancomycin, Pitocin, Insulin, Lidocaine, Heparin, Atropin, Dopamine, Epinephrine, and so on. |
KVO rate |
0.1ml/h-5.0ml/h, adjustable, choose “OFF”to cancel |
Selectable modes |
Rate mode, Time mode, Body weight mode |
Rate mode |
0.1~1500ml/h(max value depends on the syringe size) |
Body weight mode |
Body weight: 0.1~300kg Drug limit: 0.1~999.9mg Volume limit: 0.1~999.9ml Dose:0.1~9999.9mg (Subject to unit, dosage, liquid, weight and syringe) Unit: mg/kg/h, ug/kg/min |
Time mode |
Time: 1-2000min Volume limit: 0.1-999.9ml |
Preset range
|
0.1~9999.9ml |
Accumulated volume |
0.1~9999.9ml |
Alarm function |
NEAR, OVER, OCCL., syringe disengaged, no AC power supply, low battery, serious low battery, error 1/2/3, preset volume alarm |
Occlusion value |
0.04Mpa – 0.16Mpa |
Battery |
Rechargeable lithium polymer battery 7.4V, 3000mA, working for 4 hours(25ml/h) after being fully charged |
Preset time |
0-99hour, 59 minutes |
Panel instructions |
1, Infusion rate 2,yringe pump 3, Battery capacity 4, Battery charging indication 5, Accumulated injection volume 6,AC power indicator 7,BED NO. 8, Alarm information, etc |
Gross weight |
2.8kg |
Overall maximum size |
300*139*100mm (L*W*H) |
EMC |
Meet the requirements of IEC60601-1-2(2007 version) |
Classification |
Class II, type BF |
Electrical safety |
Meet the requirements of EN60601-1:1990+A1:1993+A2:1995 |
Storage condition |
Environment temperature: -20+55, air pressure (50106)kPa, relative humidity95% |
Operation condition |
Environment temperature: 5+40, air pressure (86106) kPa, relative humidity80% |
Power supply |
AC100240V 50/60HZ |
Max power consumption |
25VA |
Vet syringe pump